Uses

CardioGen-82 is a closed system used to produce rubidium Rb 82 chloride injection for intravenous administration. Rubidium Rb 82 chloride injection is indicated for Positron Emission Tomography (PET) imaging of the myocardium under rest or pharmacologic stress conditions to evaluate regional myocardial perfusion in adult patients with suspected or existing coronary artery disease.

History

There is currently no drug history available for this drug.

Other Information

11.1 Chemical Characteristics

CardioGen-82 contains accelerator-produced Sr-82 adsorbed on stannic oxide in a lead-shielded column and provides a means for obtaining sterile nonpyrogenic solutions of rubidium Rb 82 chloride injection. The chemical form of Rb-82 is 82RbCl.

The amount (millicuries) of Rb-82 obtained in each elution will depend on the potency of the generator.
When eluted at a rate of 50 mL/minute, each generator eluate at the end of elution should not contain more than 0.02 microcurie of Sr-82 and not more than 0.2 microcurie of Sr-85 per millicurie of rubidium Rb 82 chloride injection, and not more than 1 microgram of tin per mL of eluate.

11.2 Physical Characteristics

Rb-82 decays by positron emission and associated gamma emission with a physical half-life of 75 seconds.4 Table 4 shows the annihilation photons released following positron emission which are useful for detection and imaging studies.

The decay modes of Rb-82 are: 95.5% by positron emission, resulting in the production of annihilation radiation, i.e., two 511 keV gamma rays; and 4.5% by electron capture, resulting in the emission of “prompt” gamma rays of predominantly 776.5 keV. Both decay modes lead directly to the formation of stable Kr-82.4

TABLE 4
Principal Radiation Emission Data
  Mean Percent Mean Energy
Radiation Per Disintegration (keV)
Annihilation photons (2) 191.01 511 (each)
Gamma rays 13-15 776.5

The specific gamma ray constant for Rb-82 is 6.1 R/hour-millicurie at 1 centimeter. The first half-value layer is 0.7 centimeter of lead (Pb). Table 5 shows a range of values for the relative attenuation of the radiation emitted by this radionuclide that results from interposition of various thicknesses of lead.5 For example, the use of a 7.0 centimeter thickness of Pb will attenuate the radiation emitted by a factor of about 1,000.

TABLE 5
Radiation Attenuation by Lead Shielding
Shield Thickness (Pb) cm Attenuation Factor
0.7 0.5
2.3 10 -1
4.7 10-2
7.0 10-3
9.3 10-4

Sr-82 (half-life of 25 days (600 hrs)) decays to Rb-82. To correct for physical decay of Sr-82, Table 6 shows the fractions that remain at selected intervals after the time of calibration.

*Calibration time
TABLE 6
Physical Decay Chart: Sr-82 half-life 25 days
Days Fraction
Remaining
Days Fraction
Remaining
Days Fraction
Remaining
0* 1.000 15 0.660 30 0.435
1 0.973 16 0.642 31 0.423
2 0.946 17 0.624 32 0.412
3 0.920 18 0.607 33 0.401
4 0.895 19 0.591 34 0.390
5 0.871 20 0.574 35 0.379
6 0.847 21 0.559 36 0.369
7 0.824 22 0.543 37 0.359
8 0.801 23 0.529 38 0.349
9 0.779 24 0.514 39 0.339
10 0.758 25 0.500 40 0.330
11 0.737 26 0.486 41 0.321
12 0.717 27 0.473 42 0.312
13 0.697 28 0.460    
14 0.678 29 0.448    

To correct for physical decay of Rb-82, Table 1 shows the fraction of Rb-82 remaining in all 15 second intervals up to 300 seconds after time of calibration [see Dosage and Administration (2.5)].